basic specifications

Frequency characteristic

Range: 30 kHz to 220 MHz
Setting resolution: 1 mHz (30 kHz to 300 kHz) / 1 Hz (300 kHz to 220 MHz)
Accuracy: ± 1 ppm

Output characteristics

Range:
+ 8 dBm to -43 dBm (30 kHz to 300 kHz)
+10 dBm to -43 dBm (300 kHz to 1 MHz)
+18 dBm to -43 dBm (1 MHz to 200 MHz)
+10 dBm to -43 dBm (200 MHz to 220 MHz)
0.1 dBm resolution

Range setting: Start / Stop, or center / span
Sweep rate: Up to 100 μs / point
Measurement point: 1601 points
Output port: SMA (Female), 50 Ω connector

Input characteristics

Frequency range: same as signal source characteristics
Average noise level:
-70 dBm (30 kHz to 1 MHz)
-85 dBm (1 MHz to 100 MHz)
-70 dBm (100 MHz to 220 MHz)

Resolution bandwidth: 10 Hz to 15 kHz (1, 1.5, 2, 3, 4, 5, 7 steps)
Input port: SMA (Female), 50 Ω connector

Error correction function/th>

Normalize, Trans Full Cal (l Full Cal: l Open, Short, Load)

Operable PC

OS: Microsoft Windows 7 Professioanl (32 bit)
Connection port: USB 2.0 1 port

Power Supply

DC + 24 V ± 2 V / connector 39 – 29 – 1028 (manufactured by Molex)

External dimensions

260 (W) x 150 (L) x 30 (H) mm* Excluding protrusions

Operating environment

10℃ ~ 40℃
15% to 80% RH * No condensation